TOS3200
53
Setting Test Conditions of the TC Measurement
4
TC and
PCC T
est
s
Setup Items of TC2/2 (NTWK / MODE / RANGE)
Press the NEXT (SHIFT+ ) key to display the TC measurement screen 2/2 (TC2/2).
*1
Measures voltage U1 between the measurement network reference points.
Table 4-4
Determination of the measurement range
SHIFT
PREV
NEXT
Rs: 1.5 k
Ω
Cs: 0.22 μF
Rb: 0.5 k
Ω
RMS
AUTO
NTWK MODE
RANGE
NTWK
MODE
RANGE
A
A
B
Rs
Cs
Rb
TC 2 / 2
Item
Description
Panel operation
Selects the measurement network.
NTWK (F1) key
A
(For IEC 60990) (1.5 k
Ω
// 0.22 µF) + 500
Ω
NTWK
(SHIFT+F1) key
B
(For IEC 60990) (1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ 22 nF)
B1
*1
(For IEC 60065) (1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ 22 nF)
C
(For IEC 60990)
(1.5 k
Ω
// 0.22 µF) + 500
Ω
// (10 k
Ω
+ (20 k
Ω
+ 6.2 nF) // 9.1 nF)
D
1 k
Ω
E
1 k
Ω
// (10 k
Ω
+ 11.225 nF + 579
Ω
)
F
1.5 k
Ω
//0.15 µF
G
2 k
Ω
Selects the current measurement mode.
MODE (F2) key
RMS
RMS measurement
MODE
(SHIFT+F2) key
DC
DC measurement
PEAK
Peak measurement
Selects the measurement range.
RANGE (F3) key
AUTO
Automatically switches the range according to the measured
value.
RANGE
(SHIFT+F3) key
FIX
Fixes the range.
The measurement range is determined by the upper reference,
measurement network (NTWK), and current measurement mode
(MODE) settings. See Table 4-4.
NTWK
MODE
RANGE
See
Measurement
range
MODE
Upper reference
NTWK A, B, B1,C
NTWK D, E
NTWK F
NTWK G
Range 1
DC, RMS
30 µA to 600 µA
30 µA to 300 µA
30 µA to 200 µA
30 µA to 150 µA
PEAK
50 µA to 850 µA
50 µA to 424 µA
50 µA to 282 µA
50 µA to 212 µA
Range 2
DC, RMS
601 µA to 6.00 mA
301 µA to 3.00 mA
201 µA to 2.00 mA
151 µA to 1.50 mA
PEAK
851 µA to 8.50 mA
425 µA to 4.24 mA
283 µA to 2.82 mA
213 µA to 2.12 mA
Range 3
DC, RMS
6.01 mA to 30.0 mA
3.01 mA to 30.0 mA
2.01 mA to 20.0 mA 1.51 mA to 15.0 mA
PEAK
8.51 mA to 90.0 mA 4.25 mA to 45.0 mA
2.83 mA to 30.0 mA 2.13 mA to 22.5 mA
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......