TOS3200
59
Setting Additional Test Conditions
4
TC and
PCC T
est
s
1
Select the PASS HOLD on SYSTEM1/5.
2
Use the rotary knob to select the PASS HOLD time.
Measured Current Conversion Display (CONV)
The TOS3200 has the CONV function that converts the measured current using the ratio
between the measured value of the line voltage applied to the AC inlet for the EUT and the
preset CONV voltage.
For example, if the line voltage is 100.0 V and you set the preset CONV voltage to 106.0 V,
the TOS3200 will display the value equal to 106% of the measured current.
1
Select the CONV on SYSTEM1/5.
2
Use the rotary knob to set the voltage.
Buzzer Volume (BUZ VOL)
You can set the buzzer volume for PASS or FAIL judgments.
1
Select the BUZ VOL on SYSTEM1/5.
Select the BUZ VOL (PASS) to set the buzzer volume for PASS judgments, or select the
BUZ VOL (FAIL) to set the buzzer volume for FAIL judgments.
2
Use the rotary knob to set the buzzer volume (0 to 10).
You can hear the buzzer volume by pressing the BUZ CHK (SHIFT+F5) key when BUZ
VOL (PASS) or BUZ VOL (FAIL) is selected.
Time
Selectable range: 0.2 s to 10.0 s
HOLD
Hold until the STOP switch is pressed.
F5
F4
F3
F2
F1
TC
PASS
E
NC
P
E
OFF
ON
PROBE LOWER
NORM
POL
NORM
LINE
0.0W
0.0V 0.00A
NORM
COND
UPPER
ON
TIMER
0
μA
10
s
SAVE?
Yes
No
You cannot press the Yes (F4) key if
the pass hold time is short.
Voltage
Selectable range: 80.0 V to 300.0 V
OFF
Disable the conversion display.
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......