102
TOS3200
Time Settings and Calibration Management
The TOS3200 manages the calibration date using the internal system clock. If the preset cal-
ibration date is due, the message “CAL DATE EXPIRED” will appear on the screen when the
power is turned on.
Time information will be added to the stored data when you store the test data. If the system
clock is not correct, synchronize it to the present time.
1
Press the SYSTEM key to display SYSTEM1/5.
2
Press the <NEXT> (F5) key four times to display SYSTEM5/5.
Use the function key or arrow keys to move the underscore to the item you want to
change, and use the rotary knob to set the value.
The factory default settings are shown below.
TIME ADJUST (system clock)
Set the present time in the following format: year/month/day hour:minute. The time that
appears is the time when SYSTEM5/5 is displayed.
CAL. DATE (calibration date)
The date is set at the factory or by an authorized service center of Kikusui when calibration is
performed. The date is set in the following format: year/month/day. (The user cannot set this
date).
ALARM (calibration due date)
Set the calibration due date in the following format: year/month/day.
CAL DATE EXPIRED
SYSTEM
MANUAL
I / F
EDIT
AUTO
TIME ADJUST (system clock):
Standard Japanese time at factory shipment.
CAL. DATE (calibration date):
Calibration date at factory shipment.
ALARM (calibration due date):
One year after CAL. DATE.
CAL. PROTECT (calibration due pro-
tection):
OFF (disable protection).
ADJUST
ALARM
TIME ADJUST :2006/10/18 16:26
CAL. DATE
:2006/10/07
ALARM :2007/10/18
CAL.PROTECT :ON
<PREV>
SYSTEM5 / 5
Enter the present time, and press the ADJUST (F1) key to apply the entered time.
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......