UART/RS232 Triggering and Serial Decode
25
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
365
UART/RS232 Serial Decode
To set up the oscilloscope to capture UART/RS232 signals, see
To set up UART/RS232 serial decode:
1
Press
[Serial]
to display the Serial Decode Menu.
2
Press
Settings
.
3
In the UART/RS232 Settings Menu, press the
Base
softkey to select the base
(hex, binary, or ASCII) in which decoded words are displayed.
•
When displaying words in ASCII, the 7-bit ASCII format is used. Valid ASCII
characters are between 0x00 and 0x7F. To display in ASCII you must select
at least 7 bits in the Bus Configuration. If ASCII is selected and the data
exceeds 0x7F, the data is displayed in hex.
•
When
#Bits
is set to 9 in the UART/RS232 Bus Configuration Menu, the 9th
(alert) bit is displayed directly to the left of the ASCII value (which is derived
from the lower 8 bits).
4
Optional: Press the
Framing
softkey and select a value. In the decode display,
the chosen value will be displayed in light blue. However, if a parity error occurs
the data will be displayed in red.
5
If the decode line does not appear on the display, press the
[Serial]
key to turn it
on.
NOTE
For UART/RS232 triggering setup see
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...