Digital Voltmeter
16
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
243
•
Frequency
— displays the frequency counter measurement.
6
Press
Transparent
to toggle between a transparent and shaded background for
the DVM display.
7
If the selected source channel is not used in oscilloscope triggering, press
Auto
Range
to disable or enable automatic adjustment of the DVM channel's vertical
scale, vertical (ground level) position, and trigger (threshold voltage) level (used
for the counter frequency measurement).
When enabled,
Auto Range
overrides attempted adjustments of the channel's
vertical scale and position knobs.
When disabled, you can use the channel's vertical scale and position knobs
normally.
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...