Utility Settings
20
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
277
To establish a LAN connection
Automatic
Configuration
1
Press
[Utility] > I/O
.
2
Press the
LAN Settings
softkey.
3
Press the
Config
softkey; then, turn the Entry knob to select
Automatic
, and press
the softkey again to enable it.
If your network supports DHCP or AutoIP, enabling
Automatic
lets the
oscilloscope use those services to get its LAN configuration settings
4
If your network provides Dynamic DNS, you can enable the
Dynamic DNS
option
to let the oscilloscope register its hostname and use the DNS server for name
resolution.
5
You can enable the
Multicast DNS
option to let the oscilloscope use Multicast
DNS for name resolution on small networks without a conventional DNS server.
6
Connect the oscilloscope to the local area network (LAN) by inserting the LAN
cable into the "LAN" port on the rear panel of the oscilloscope.
In a few moments the oscilloscope will connect to the network automatically.
If the oscilloscope does not automatically connect to the network, press
[Utility]
> I/O > LAN Reset
. In a few moments the oscilloscope will connect to the
network.
Manual
Configuration
1
Get the oscilloscope's network parameters (host name, IP address, subnet
mask, gateway IP, DNS IP, etc.) from your network administrator.
2
Press
[Utility] > I/O
.
3
Press the
LAN Settings
softkey.
4
Press the
Config
softkey; then, turn the Entry knob to select
Automatic
, and press
the softkey again to disable it.
NOTE
When you connect the oscilloscope to a LAN it is a good practice to limit access to the
oscilloscope by setting a password. By default, the oscilloscope is not password protected.
See
on page 305 to set a password.
NOTE
Any time you modify the oscilloscope's hostname, it breaks the connection between the
oscilloscope and the LAN. You need to re-establish communication to the oscilloscope using
the new hostname.
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...