Digital Channels
6
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
109
3
Press the
D7 - D0
softkey, then select a logic family preset or select
User
to
define your own threshold.
The threshold you set applies to all channels within the selected D7 - D0 group.
Each of the two channel groups can be set to a different threshold if desired.
Values greater than the set threshold are high (1) and values less than the set
threshold are low (0).
If the
Thresholds
softkey is set to
User
, press the
User
softkey for the channel group,
then turn the Entry knob to set the logic threshold. There is one
User
softkey for
each group of channels.
To reposition a digital channel
1
Make sure the multiplexed scale and position knobs to the right of the key are
selected for digital channels.
If the arrow to the left of the
[Digital]
key is not illuminated, press the key.
2
Use the multiplexed Select knob to select the channel.
The selected waveform is highlighted in red.
3
Use the multiplexed Position knob to move the selected channel waveform.
If a channel waveform is repositioned over another channel waveform, the
indicator at the left edge of the trace will change from
D
nn designation (where
nn is a one or two digit channel number) to
D
*. The "*" indicates that two
channels are overlaid.
Logic family
Threshold Voltage
TTL
+1.4 V
CMOS
+2.5 V
ECL
–1.3 V
User
Variable from –8 V to +8 V
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...