Utility Settings
20
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
285
•
Mask
— The pass/fail status is evaluated periodically. When the evaluation of
the testing period results in a failure, the trigger output pulses high (+5 V).
Otherwise, the trigger output remains at low (0 V). See
Testing,” starting on page 227.
•
Waveform Generator Sync Pulse
— All of the waveform generator output
functions (except DC and Noise) have an associated Sync signal:
The Sync signal is a TTL positive pulse that occurs when the waveform rises
above zero volts (or the DC offset value).
, “Waveform Generator,” starting on page 245.
The TRIG OUT connector also provides the User Cal signal. See
Enabling Remote Command Logging
When remote command logging is enabled, remote commands sent to the
instrument (and results returned by the instrument) can be logged to the screen,
to a text file on a USB storage device, or to both the screen and a text file.
To enable remote command logging:
1
Press
[Utility] > Options > Remote Log
to open the Remote Log menu:
2
Press
Enable
to enable or disable the remote command logging feature.
When remote logging is enabled, additional debug information can be included
in the returned error string. If the error is detected by the SCPI command
parser, such as a header error or other syntax error, the extra debug
information is generated and included. But if the error is detected by the
oscilloscope system, such as when an out-of-range value is sent, then no extra
debug information is included.
3
Press
Destination
to select whether remote commands are logged to a text file
(on a connected USB storage device), logged to the screen, or both.
4
Press
Write Mode
to specify whether logged commands will be created in a new
list or appended to existing logged commands.
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...