330
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
23
CAN/LIN Triggering and Serial Decode
3
Press the
Trigger:
softkey; then, turn the Entry knob to select the trigger
condition:
•
SOF - Start of Frame
— The oscilloscope triggers at the start of a frame.
•
Remote Frame ID (RTR)
— The oscilloscope triggers on remote frames with the
specified ID. Press the
Bits
softkey to select the ID.
•
Data Frame ID (~RTR)
— The oscilloscope will trigger on data frames matching
the specified ID. Press the
Bits
softkey to select the ID.
•
Remote or Data Frame ID
— The oscilloscope will trigger on remote or data
frames matching the specified ID. Press the
Bits
softkey to select the ID.
•
Data Frame ID and Data
— The oscilloscope will trigger on data frames
matching the specified ID and data. Press the
Bits
softkey to select the ID
and set up the number of data bytes and values.
•
Error Frame
— The oscilloscope will trigger on CAN active error frames.
•
All Errors
— The oscilloscope will trigger when any form error or active error is
encountered.
•
Acknowledge Error
— The oscilloscope will trigger when the acknowledge bit is
recessive (high).
•
Overload Frame
— The oscilloscope will trigger on CAN overload frames.
4
If you select a condition that lets you trigger on ID or data values, use the
Bits
softkey and the CAN Bits Menu to specify those values.
For details about using the CAN Bits Menu softkeys, press and hold the softkey
in question to display the built-in help.
You can use the
Zoom
mode for easier navigation of the decoded data.
NOTE
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the
oscilloscope is AutoTriggering. Press the
[Mode/Coupling]
key, then press the
Mode
softkey
to set the trigger mode from
Auto
to
Normal
.
NOTE
To display CAN serial decode, see
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...