Mask Testing
15
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
233
To manually modify a mask file
You can manually modify a mask file that you created using the Automask
function.
1
Follow the steps 1-7 in
"To create a mask from a "golden" waveform
on page 227. Do not clear the mask after creating it.
2
Attach a USB mass storage device to the oscilloscope.
3
Press the
[Save/Recall]
key.
4
Press the
Save
softkey.
5
Press the
Format
softkey and select
Mask
.
6
Press the second softkey and select a destination folder on your USB mass
storage device.
7
Press the
Press to Save
softkey. This creates an ASCII text file that describes the
mask.
8
Remove the USB mass storage device and connect it to a PC.
9
Open the .msk file your created using a text editor (such as Wordpad).
10
Edit, save, and close the file.
The mask file contains the following sections:
• Mask File Identifier.
• Mask Title.
Reset Statistics
Note that statistics are also reset when:
•
Mask Test is switched on after being switched off.
•
Clear Mask softkey is pressed.
•
An Automask is created.
Additionally, the accumulated time counter is reset whenever the oscilloscope
is run after the acquisition was stopped.
Transparent
Enable Transparent mode to write measurement values and statistics on the
screen without a background. Disable the Transparent mode to show them with
a gray background. The Transparent setting affects mask test statistics,
measurement statistics, and reference waveform information display.
Clear Display
Clears acquisition data from the oscilloscope display.
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...