224
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
14
Measurements
You can also open the Measurement Threshold Menu by pressing
[Analyze] >
Features
and then selecting
Measurement Thresholds
.
2
Press the
Source
softkey to select the analog channel or math waveform source
for which you want to change measurement thresholds.
Each analog channel and the math waveform can be assigned unique threshold
values.
3
Press the
Type
softkey to set the measurement threshold to
%
(percentage of
Top and Base value) or to
Absolute
(absolute value).
•
Percentage thresholds can be set from 5% to 95%.
•
The units for absolute threshold for each channel is set in the channel probe
menu.
•
When the
Source
is set to
Math: f(t)
, the threshold
Type
can only be set to
Percent
.
4
Press the
Lower
softkey; then, turn the Entry knob to set the lower measurement
threshold value.
Increasing the lower value beyond the set middle value will automatically
increase the middle value to be more than the lower value. The default lower
threshold is 10% or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can be set from 5% to
93%.
5
Press the
Middle
softkey; then, turn the Entry knob to set the middle
measurement threshold value.
TIP
Absolute threshold hints
•
Absolute thresholds are dependent on channel scaling, probe attenuation, and probe units. Always
set these values first before setting absolute thresholds.
•
The minimum and maximum threshold values are limited to on-screen values.
•
If any of the absolute threshold values are above or below the minimum or maximum waveform
values, the measurement may not be valid.
Содержание DSO-X 2014A
Страница 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Страница 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Страница 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Страница 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Страница 268: ...268 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...
Страница 370: ...370 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 25 UART RS232 Triggering and Serial Decode...