BHK-MG 1/2 (OPR) 030916
B-15
FIGURE B-6. USING STATUS COMMANDS AND QUERIES
B.61
STATus:OPERation:ENABle? QUERY
STAT:OPER:ENAB?
Syntax
:
Short Form: STAT:OPER:ENAB?
Long Form: STATus:OPERation:ENABle?
Return Value: <int_value> 0 to 1313 (1 + 32 + 256 + 1024).
Description:
Reads Operation Enable Register (see Table B-2).
Returns value of Operation Enable Register bits.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-6.)
B.62
STATus:OPERation[:EVENt] QUERY
STAT:OPER?
Syntax
:
Short Form: STAT:OPER[:EVEN]?
Long Form: STATus:OPERation[:EVENt]?
Return Value: <int_value> 0 to 1313 (1 + 32 + 256 + 1024).
Description:
Indicates changes in conditions monitored by Operational Event Register.
Returns the value of
the Operation Event register. The Operation Event register is a read-only register which holds
(latches) all events that occur. Reading the Operation Event register clears it. (See example, Figure B-
6.)
B.63
STATus:PRESet COMMAND
STAT:PRES
Syntax
:
Short Form: STAT:PRES
Long Form: STATus:PRESet
Description:
Presets the Status Questionable Enable register.
This command sets all bits of the Operation
Enable (PAR. B.60) and Questionable Enable (Table B-3) Registers to 0, preventing all status events
from being reported. (See example, Figure B-6.)
NOTE:The power supply is assumed to be operating in CV (constant voltage) mode.
OUTP ON
Output Enabled.
VOLT 3;CURR 1E-2
Program output to 3V, 0.01A
STAT:OPER:ENAB 1056
Mask enabled for CC, WTG and bits.
STAT:OPER:ENAB?
Returns 1056 (32 + 1024) (CC, WTG bits set).
STAT:QUES:ENAB 3
Mask enabled for OV and OC bits (1 + 2).
STAT:QUES:ENAB?
Returns 3 (1 + 2) indicating OV and OC bits are enabled.
STAT:PRES
Operation Condition Enable and Questionable Condition Enable
registers are preset.
INIT:CONT ON
Continuous triggers enabled.
STAT:OPER:COND?
Power supply returns 288 (256 + 32) to indicate that power
supply is constant voltage mode and Wait For Trigger is true.
--- OUTPUT MOMENTARILY SHORTED.
STAT:OPER?
Returns 1312, e.g., indicating that since the last reading of the
Operation Event Register the power supply has entered
Constant Current mode, the Wait Trigger was set.
STAT:OPER?
Returns 0 indicating no changes since previous reading of the
Operation Event register.
STAT:QUES?
Returns 0 (no questionable conditions occurred since previous
reading
--- OVERTEMPERATURE CONDITION OCCURS.
STAT:QUES?
Returns 8 (overtemperature condition tripped since the last
STAT:QUES? query).
STAT:QUES:COND?
Returns 8, (Power supply still in overtemperature condition).
STAT:QUES?
Returns 0, (Power supply no longer in overtemperature
condition).
STAT:QUES:COND?
Returns 8, (Power supply still in overtemperature state).
SYST:ERR?
Power supply returns 0,“No error” message.
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......