BHK-MG 1/2 (OPR) 030916
B-13
B.52
[SOURce:]VOLTage:MODE COMMAND
VOLT:MODE
Syntax:
Short Form: [SOUR:]VOLT:MODE (FIX | LIST)
Long Form: [SOURce:]VOLTage:MODE (FIXed | LIST)
Description:
Allows the user to execute or stop a list.
The default mode is FIX: the power supply executes com-
mands normally, and LIST commands can be issued to establish the parameters and sequence of a
list.
When VOLT:MODE LIST is issued, a list is executed (See LIST commands and Figure B-3). While
the list is being executed, LIST commands are not accepted and will produce a command error.
Issuing VOLT:MODE FIX while the list is running will stop the execution of the list and return power
supply to settings in effect prior to running the list. If the list runs to completion, the settings of the last
step of the list will be in effect.
Related Commands: LIST commands. (See example, Figure B-3.)
B.53
[SOURce:]VOLTage:MODE? QUERY
VOLT:MODE?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:MODE?
Long Form: [SOURce:]VOLTage[:LEVel]:MODE?
Return value:
FIXED or LIST
Description:
Identifies active voltage mode.
Returns LIST while list is being executed. Returns FIXED while in
fixed (default) mode of operation. Related Commands: LIST commands. (See example, Figure B-3.)
B.54
[SOURce:]VOLTage:PROTection[:LEVel] COMMAND
VOLT:PROT
Syntax:
Short Form: [SOUR:]VOLT:PROT[:LEV] <exp_value>
Long Form: [SOURce:]VOLTage:PROTection[:LEVel] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets overvoltage protection level for power supply.
If the value is out of the acceptable overvoltage
range for the power supply model, error message -222,”Data out of range” is posted in the output
queue. If the power supply output exceeds the voltage protection level programmed, the power supply
circuit breaker trips. (See example, Figure B-5.)
B.55
[SOURce:]VOLTage:PROTection[:LEVel]? QUERY
VOLT:PROT?
Syntax:
Short Form: [SOUR:]VOLT:PROT[:LEV] MIN, MAX
Long Form: [SOURce:]VOLTage:PROTection[:LEVel] MIN, MAX
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Identifies overvoltage protection setting, maximum allowable overvoltage protection, or mini-
mum overvoltage protection.
VOLT:PROT? returns value set by VOLT:PROT. VOLT:PROT? MAX
returns maximum voltage protection value (see Table 1-2); this value is determined at the factory and
cannot be changed by the user. VOLT:PROT? MIN returns the minimum voltage protection value
(always 0). (See example, Figure B-5.)
B.56
[SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude] COMMAND
VOLT:TRIG
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP] <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Programs voltage value to be transferred to output by *TRG commands.
Actual output voltage
will depend on load conditions. If the value exceeds the VOLT:LIM[:HIGH] value, error message -
222,”Data out of range” is posted in output queue. (See example, Figure B-1.)
B.57
[SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude]? QUERY
VOLT:TRIG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Returns value representing voltage value to be programmed by *TRG command established by
VOLT:TRIG command).
(See example, Figure B-1.)
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......