2-6
BHK-1/2-MG SERIES 030916
2.3
SOURCE POWER REQUIREMENTS
BEFORE APPLYING AC SOURCE POWER TO THE POWER SUPPLY, VERIFY THAT THE
LINE VOLTAGE TO BE SUPPLIED MATCHES THE POSITION OF THE AC INPUT SELEC-
TOR SWITCH AT THE REAR PANEL (FACTORY DEFAULT IS 115V).
This power supply operates with the installed circuit breaker from single phase AC mains power
over the specified voltage and frequency ranges without adjustment or modification. The AC
INPUT SELECTOR switch located on the rear panel (Figure 2-2) allows selection of either
115Vac or 230Vac source power.
The maximum AC current absorbed by the BHK-MG 40W Power Supply from the power source
is 1.1A a-c @115V a-c and 0.6A a-c @230V a-c; the power source must be able to deliver the
maximum current as well as surge current ten times larger than maximum.
2.4
COOLING
The power devices used within the power supply are maintained within their operating tempera-
ture range by means of internal heat sink assemblies cooled by two internal (d-c type) cooling
fans.
ALL INLET AND EXHAUST OPENINGS AROUND THE POWER SUPPLY CASE MUST BE
KEPT CLEAR OF OBSTRUCTION TO ENSURE PROPER AIR ENTRY AND EXHAUST.
Periodic cleaning of the power supply interior is recommended. If the power supply is rack
mounted, or installed within a confined space, care must be taken that the ambient temperature,
which is the temperature of the air immediately surrounding the power supply, does not rise
above the specified limits (see Table 1-2).
TABLE 2-8. CURRENT PROGRAMMING TERMINAL STRIP A6TB3, TERMINAL ASSIGNMENTS
TERMINAL
SIGNAL NAME
FUNCTION
1
–10V
–10V d-c reference voltage
2
C(NINV)
Noninverting input of uncommitted amplifier
3
SGND
Signal common
4
C(+IN)
Programming input for positive input signal
5
+10V
+10V d-c reference voltage
6
C(INV)
Inverting input of uncommitted amplifier
7
C(FBK)
Internal feedback resistor; the other end is connected to uncommitted
amplifier output
8
C(OUT)
Output of uncommitted amplifier
9
C EXT
External analog programming voltage input: 0 to –10V programs 0 to
100% of IOmax.
10
SGND
Signal common
WARNING
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......