BHK-MG 1/2 OPR 030916
4-7
4.5.4
CALIBRATION PROCEDURE
1. Click on the
Calibrate
button on the main control panel (see Figure 4-2). This opens a cali-
bration panel (see Figure 4-3 and Table 4-2) that allows calibration in either voltage or cur-
rent mode.
2. After initializing, the password window appears. The password has been set at the factory
(see Table 4-3). Type the proper password in the text box and click
OK
. If the password has
been forgotten, consult factory. If the password is correct,.the calibration panel (Figure 4-3) is
active and calibration can be accomplished by following the screen prompts.
FIGURE 4-4. PASSWORD ENTRY
3. For Voltage Calibration, click
YES
button (see Figure 4-5).
FIGURE 4-5. VOLTAGE CALIBRATION
4. The text window reads
Connect DVM to Output Sense, Remove Shunt
. Remove
any connections bOUT and –OUT. Connect DVM (high) to +OUT and DVM (low)
to –OUT, then click
OK
button.
NOTE: Use external DVM for all measurements.
5. The text window reads
Vout ZERO adjust
. Using the
<
and
>
buttons (see PAR. 4.5.3),
adjust the output voltage until the DVM is as close to 0 as possible but slightly positive, then
click
OK
button.
6. The text window reads
Vout F.S.
Using the coarse
<<
and
>>
and fine
<
and
>
buttons
(see PAR. 4.5.3), adjust the output voltage until the DVM is as close as possible, but above,
the rated nominal voltage, e.g., 1000.00V for the BHK 1000-40MG, then click
OK
button.
7. The text window reads
Cal Current ? no=skip.
Click
YES
button to perform current
calibration.
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......