4. Enter 4-digit password and press
ENTER
. Factory calibration values are restored.
4.9
CALIBRATION STORAGE
The BHK-MG 40W Power Supply. maintains the calibration tables in Flash Memory until a PACK
is executed. There are six calibration areas maintained in Flash Memory: Working, Prior, Oldest,
Factory, Master, and First. The calibration password is required In order to store the calibration.
The calibration can be copied to another area using the CAL:DUMP? and CAL:COPY com-
mand. The syntax is as follows: CAL:DUMP? [source];:CAL:COPY [destination], where [source]
and [destination] refer to the areas of Flash memory where calibration data is stored, designated
as: WORKing, PRIor, OLDest, FACTory, MASTer, FIRst. Source refers to the calibration area
that is to be copied, destination to the area that the calibration will be copied into.
The ability to save to the FACTORY area is prohibited. If a user needs to overwrite the factory
calibration, a special firmware needs to be created to provide a special password for this capa-
bility. Each customer that requests this capability will be provided a password that the company
can specify. it must contain at least one alpha, one numeric and one punctuation character
(!@#$%^&*) and be at least 6 characters long and not longer than 20 characters in length. All
alpha data must be upper case and no control nor space characters are allowed.
The Master calibration should never be overwritten. Factory, Master, and First are set to the
same values when the power supply is calibrated at the factory. The Working calibration is the
active calibration. Each time a CAL:SAV is executed, the values are saved in the Working
(active) area. At the same time, the values previously stored in Working are moved to Prior, and
the values previously stored in Prior are moved to Oldest. The values previously stored in Old-
est are no longer available. Table 4-4 illustrates calibration storage and the use of the
CAL:DUMP? query and CAL:COPY command.
TABLE 4-4. CALIBRATION STORAGE
COMMAND
WORKing
PRIor
OLDest
FACTory
MASTer
FIRst
1. CAL:SAVE
Cal 1 values
Factory cal.
Master cal.
First cal.
2. CAL:SAVE
Cal 2 values
Cal 1 values
No Change
No Change
No Change
3. CAL:SAVE
Cal 3 values
Cal 2 values
Cal 1 values
No Change
No Change
No Change
4. CAL:SAVE
Cal 4 values
Cal 3 values
Cal 2 values
No Change
No Change
No Change
5. CAL:DUMP? FACTory;:CAL:COPY WORKing
Factory cal.
Cal 4 values
Cal 3 values
No Change
No Change
No Change
6. CAL:DUMP? PRIor;:CAL:COPY WORKing
Cal 4 values
Factory cal.
Cal 4 values
No Change
No Change
No Change
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......