3-8
BHK-1/2-MG 030916
3.2.9
SETTING OUTPUT VOLTAGE OR CURRENT
V SET
and
I SET
set output voltage and current limit, respectively, when the unit is in constant
voltage (CV) mode and set voltage limit and output current, respectively, when the unit is in con-
stant current (CC) mode. The mode (CV or CC) is determined by the load together with the pro-
grammed settings. As long as the voltage across the load produces a current that is less than
the
I SET
value, the unit operates in CV mode (voltage programmed to
V SET
value, current lim-
ited by
I SET
value). If the load changes to the point that current through the load reaches the
I
SET
value, the unit automatically enters CC mode (current programmed to
I SET
value, voltage
limited by
V SET
value).
With the power supply in command entry status
(:_:_:)
, press
V SET
(voltage) or
I SET
(cur-
rent) key to put the power supply in data entry status
(=_=_=)
. The lower right side of the LCD
shows the programmed setting in effect when the key is pressed. Press
ENTER
or
CLEAR
to
exit without changing setting. Enter new value (the
key erases data just entered) and press
ENTER
to accept new setting or
CLEAR
to exit without changing setting.
NOTE: Two decimal places are used for voltage settings and three decimal places for current
settings (except that only one decimal place is used for setting voltage on Models BHK
1000-40MG and BHK 2000-20MG).
With the power supply in command entry status
(:_:_:)
, an alternative is to use
key
(increase) or
key (decrease) to change the output by the minimum increment (0.025% of
E
O
max for voltage, 0.025% of I
O
max for current). In CC mode, these keys control output cur-
rent; in CV mode they control output voltage. When the unit switches from constant current (CC)
to constant voltage (CV) mode, or from CV to CC mode the
and
keys are reassigned to
the active parameter. Once the transition point is reached, each press of the
key increases
voltage, then current, changing the mode, from CV to CC or from CC to CV. The
key has no
effect after the maximum (or programmed limit) has been reached.
If the value entered exceeds the maximum setting, the LCD bottom line flashes
VsetMAX=n.n
V
or
IsetMAX=n.n mA
where n.n is the maximum setting (see PAR.3.2.11). Press
CLEAR
and
enter a value that does not exceed the maximum setting.
3.2.10
SETTING OVERVOLTAGE OR OVERCURRENT PROTECTION
With the power supply in command entry status
(:_:_:)
, press
OV SET
(overvoltage) or
OC
SET
(overcurrent) to put the power supply in data entry status
(=_=_=)
. The lower right side of
the LCD shows the programmed setting in effect when the key is pressed. Press
ENTER
or
CLEAR
to exit without changing setting. Enter new value and press
ENTER
to accept new set-
ting (the
key erases data just entered), or
CLEAR
to exit without changing setting.
The value for overvoltage protection can be set within the range of 0 to 1.1 x E
O
max; overcur-
rent can be set within the range of 0 to 1.1 x I
O
max. The factory default values are 1.1 x E
O
max
for overvoltage protection and 1.1 x I
O
max for overcurrent protection.
NOTE: Two decimal places are used for voltage settings and three decimal places for current
settings (except that only one decimal place is used for setting voltage on Models BHK
1000-40MG and BHK 2000-20MG).
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......