B-20
BHK-MG 1/2 (OPR) 030916
B.89
SYSTem:VERSion? QUERY
SYST:VERS?
Syntax
:
Short Form: SYST:VERS?
Long Form: SYSTem:VERSion?
Return Value: <int_value>.<int_value> (YYYY.V)
Description:
Identifies SCPI Version implemented.
Returns SCPI Version number:
YYYY = year, V = Revision number for specified year. (See example, Figure B-7.
B.90
TRIGger:SOURce COMMAND
TRIG:SOUR
Syntax
:
Short Form: TRIG:SOUR {EXT | BUS}
Long Form: TRIGger:SOURce {EXT | BUS}
Description: Sending TRIG:SOUR activates either internal (BUS) or external (EXT) trigger. When the external trig-
ger (ground) is applied to J2 pin 3, output voltage and current are set to previously stored values for
VOLT:TRIG and CURR:TRIG.
NOTE: If INIT:CONT 1 was sent prior to sending TRIG:SOUR EXT, continuous mode is disabled and
unit will wait for INIT:IMM
TABLE B-4. ERROR MESSAGES
ERROR MESSAGE
ESR ERROR BIT SET
(SEE PAR. A.5)
EXPLANATION
0,“No error”
None
No error
-100,“Command error”
Command Error bit 5
Command and data understood, but more information included which is
not recognized.
-102,”Syntax error”
Command Error bit 5
First 4 characters recognized, subsequent characters not recognized.
-103,”Invalid separator”
Command Error bit 5
For example, VOLT.PROT received instead of VOLT:PROT
-108,”Parameter Not
Allowed Error”
Command Error bit 5
Volt12 sequence, channel number is invalid
-109,”Missing
parameter”
Command Error bit 5
For example, VOLT instead of VOLT 21.
-111,”Header
separator error”
Command Error bit 5
Missing space between volt and value or ; missing
-113,”Undefined
header”
Command Error bit 5
First 4 characters could not be identified as legal command.For exam-
ple, command VLT instead of VOLT
-120,”Numeric data
error”
Command Error bit 5
Expected number but other characters were detected
-121,”Invalid character
in number”
Command Error bit 5
Volt 1,500 (comma not allowed)
-123,”Exponent too
large”
Command Error bit 5
Exponent E+3 or greater is invalid.
-141,”Invalid character
data”
Command Error bit 5
For example OUTP OFD or OUTP STOP instead of OUTP OFF
-150,”String data
error”
Command Error bit 5
Invalid characters were detected in numeric entry.For example E.1
instead of E+1 or 4d3 instead of 4.3.
-203,”Command
protected”
Execution error bit 4
Command only valid in password Cenable state
-221,”Settings
Conflict”
Execution error bit 4
Invalid password from syst:pass:cen command
-222,“Data out of
range”
Execution error bit 4
Value exceeds power supply rating
Содержание BHK-MG Series
Страница 2: ......
Страница 6: ......
Страница 16: ......
Страница 18: ...xii BHK MG030916 FIGURE 1 1 BHK MG 40W SERIES PROGRAMMABLE POWER SUPPLY...
Страница 24: ...1 6 BHK 1 2 MG OPR 030916 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Страница 25: ...BHK 1 2 MG OPR 030916 1 7 FIGURE 1 2 BHK 40W SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 26: ...1 8 BHK 1 2 MG OPR 030916 FIGURE 1 3 RA 24 RACK ADAPTER WITH TWO 1 2 RACK BHK MG 40W UNITS...
Страница 35: ...BHK 1 2 MG SERIES 030916 2 3 FIGURE 2 2 BHK MG SERIES REAR PANEL CONTROLS AND CONNECTIONS...
Страница 100: ...3 54 BHK 1 2 MG 030916 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION...
Страница 108: ...3 62 BHK 1 2 MG 030916 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION Eo is the output voltage of the power assembly...
Страница 113: ...BHK 1 2 MG 030916 3 67 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION...
Страница 150: ......