Series 3700 System Switch/Multimeter Reference Manual
Section 11
:
Calibration
Document Number: 3700S-901-01 Rev. A / August 2007
11-9
DC current calibration
Make the connections as shown, then perform the following calibration steps (DC Cal Step 10
through Step 14):
Figure 11-4: DC current verification 10uA-100uA ranges
DC Cal Step 10: 100uA
1. Send the following commands:
SEND dmm.func = dmm.DC_CURRENT
SEND dmm.range = 100e-6
2. Source 100µA.
3. Send the following commands:
SEND dmm.calibration.dc(10,.0001) print("done")