Section 9
:
Instrument Control Library (ICL)
Series 3700 System Switch/Multimeter Reference Manual
9-172
Document Number: 3700S-901-01 Rev. A / August 2007
scan.trigger.measure.stimulus
Remarks
This attribute selects which event(s) will cause the measure event detector to enter the
detected state. Set this attribute to nil to bypass waiting for an event.
eventid
may be one of the following (existing trigger event ID's):
digio.trigger[line].EVENT_ID: An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
display.trigger.EVENT_ID: The trigger key on the front panel is pressed.
trigger.EVENT_ID: A *trg message on the active command interface. If GPIB is
the active command interface, a GET message will also generate this event. Also,
the trigger.set function will generate this event.
trigger.blender[N].EVENT_ID: A combination of events has occurred.
trigger.timer[N].EVENT_ID: A delay expired. These can be expanded to include
LXI alarms.
tsplink.trigger[N].EVENT_ID: An edge (either rising, falling, or either based on the
configuration of the line) on the tsplink trigger line.
lan.trigger[N].EVENT_ID
scan.trigger.EVENT_SCAN_READY: Scan Ready Event.
scan.trigger.EVENT_SCAN_START: Scan Start Event
scan.trigger.EVENT_CHANNEL_READY: Channel Ready Event
scan.trigger.EVENT_MEASURE_COMP: Measure Complete Event
scan.trigger.EVENT_SEQUENCE_COMP: Sequence Complete Event
scan.trigger.EVENT_SCAN_COMP: Scan Complete Event
scan.trigger.EVENT_IDLE: Idle Event
Use this to start a set of measure count readings that are being triggered by a single
event. If desire an event to pace the readings then, use the sample trigger source.
Also see
scan.trigger.sequence.stimulus
(on page 9-173)
Example
To set trigger stimulus of the measure event detector to channel ready event:
scan.trigger.measure.stimulus =
scan.trigger.EVENT_CHAN_READY
scan.trigger.sequence.clear
Function
Clear the sequence event detector.
Usage
scan.trigger.sequence.clear()
Remarks
This function will set the sequence event detector to the undetected state.
Example
To clear the sequence event detector:
scan.trigger.sequence.clear()
scan.trigger.sequence.set
Function
Set the sequence event detector to the detected state.
Usage
scan.trigger.sequence.set()