Section 9
:
Instrument Control Library (ICL)
Series 3700 System Switch/Multimeter Reference Manual
9-190
Document Number: 3700S-901-01 Rev. A / August 2007
status.node_enable
Attribute
Status node enable register.
Reads status node enable register:
nodeenabreg = status.node_enable
Writes to system enable register:
status.node_enable = nodeenabreg
Set
nodeenabreg
to one of the following values:
To clear all bits:
0
To set (enables) MSB bit (B0):
status.MEASUREMENT_SUMMARY_BIT
- or -
status.MSB
To set (enables) EAV bit (B2):
status.ERROR_AVAILABLE
- or -
status.EAV
To set (enables) QSB bit (B3):
status.QUESTIONABLE_SUMMARY_BIT
- or -
status.QSB
To set (enables) MAV bit (B4):
status.MESSAGE_AVAILABLE
- or -
status.MAV
To set (enables) ESB bit (B5):
status.EVENT_SUMMARY_BIT
- or -
status.ESB
To set (enables) MSS bit (B6):
status.MASTER_SUMMARY_STATUS
- or -
status.MSS
To set (enables) OSB bit (B7):
status.OPERATION_SUMMARY_BIT
- or -
status.OSB