Section 9
:
Instrument Control Library (ICL)
Series 3700 System Switch/Multimeter Reference Manual
9-206
Document Number: 3700S-901-01 Rev. A / August 2007
status.system4.*
.condition
.enable
.event
Remarks
In an expanded system (TSP- Link), this attribute is used to read or write to the
system4 node registers.
Reading a system4 node register returns a numeric value whose binary equivalent
indicates which register bits are set. The bits of the system4 node register are
identified as follows:
B0-EXT bit
B4-Node 46
B8-Node 50
B12-Node 54
B1-Node43
B5-Node47
B9-Node51
B13-Node55
B2-Node44
B6-Node48
B10-Node52
B14-Node56
B3-Node45
B7-Node49
B11-Node53
B15-Not used
For example, assume value 9 is returned for the enable register. The binary
equivalent is 0000000000001001. This value indicates that bit 0 (EXT) and bit 3
(Node 45) are set.
Assigning a value to the status.system4.enable attribute sets the extension bit or a
node bit of the system4 node enable register.
Also see
status.system.*
(on page 9-201)
status.system2.*
(on page 9-203)
status.system3.*
(on page 9-204)
status.system5.*
(on page 9-207)
Example
Sets the extension bit of the system4 enable register:
status.system4.enable = status.system4.EXT