Section 10
:
Verification
Series 3700 System Switch/Multimeter Reference Manual
10-6
Document Number: 3700S-901-01 Rev. A / August 2007
6. Source positive and negative full-scale and half-scale voltages for each of the ranges listed
below. For each voltage setting, be sure that the reading is within stated limits.
Figure 10-1: DC volts verification
DCV verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
Example reading limit calculation
(on page 10-3)).
Applied voltage
Range
Voltage
Lower limit
Upper limit
REL U.U.T. for 100mV
1.00E-01
0V
Verify DCV 100mV
1.00E-01
1.00E-01
9.999610E-02
1.000039E-01
Verify DCV 100mV
1.00E-01
5.00E-02
4.999760E-02
5.000240E-02
Verify DCV 100mV
1.00E-01
-5.00E-02
-5.000240E-02
-4.999760E-02
Verify DCV 100mV
1.00E-01
-1.00E-01
-1.000039E-01
-9.999610E-02
REL U.U.T. again for 1V
1.00E+00
0V
Verify DCV 1V
1.00E+00
1.00E+00
9.999680E-01
1.00
Verify DCV 1V
1.00E+00
5.00E-01
4.999830E-01
5.000170E-01
Verify DCV 1V
1.00E+00
-5.00E-01
-5.000170E-01
-4.999830E-01
Verify DCV 1V
1.00E+00
-1.00E+00 -1.00
-9.999680E-01
Disable REL
Verify DCV 10V
1.00E+01
1.00E+01
9.00
1.01
Verify DCV 10V
1.00E+01
5.00E+00
4.00
5.00
Verify DCV 10V
1.00E+01
0.00E+00
-2.000000E-05
2.000000E-05
Verify DCV 10V
1.00E+01
-5.00E+00 -5.00
-4.00
Verify DCV 10V
1.00E+01
-1.00E+01 -1.01
-9.00
Disable REL