Section 9
:
Instrument Control Library (ICL)
Series 3700 System Switch/Multimeter Reference Manual
9-74
Document Number: 3700S-901-01 Rev. A / August 2007
dmm.calibration.ac
Remarks
Use this command to indicate the desired AC calibration step to perform on the DMM.
This command will generate an error if the step is out of sequence, does not exist, or
the calibration is locked. An error will be generated if the calibration step does not
complete successfully or if the value passed is invalid for the step, out of range, or not
needed.
Details
AC volts calibration
dmm.calibration.ac(1) ' AC cal step 1
(10mV, 1kHz step)
dmm.calibration.ac(2) ' AC cal step 2
(100mV, 1kHz step)
dmm.calibration.ac(3) ' AC cal step 3
(100mV, 50kHz step)
dmm.calibration.ac(4) ' AC cal step 4
(1V, 1kHz step)
dmm.calibration.ac(5) ' AC cal step 5
(1V, 50kHz step)
dmm.calibration.ac(6) ' AC cal step 6
(10V, 1kHz step)
dmm.calibration.ac(7) ' AC cal step 7
(10V, 50kHz step)
dmm.calibration.ac(8) ' AC cal step 8
(100V, 1kHz step)
dmm.calibration.ac(9) ' AC cal step 9
(100V, 50kHz step)
dmm.calibration.ac(10) ' AC cal step 10
(300V, 1kHz step)
AC current calibration
dmm.calibration.ac(11) ' AC cal step 11
(100uA 1kHz step)
dmm.calibration.ac(12) ' AC cal step 12
(1mA 1kHz step)
dmm.calibration.ac(13) ' AC cal step 13
(10mA 1kHz step)
dmm.calibration.ac(14) ' AC cal step 14
(100mA 1kHz step)
dmm.calibration.ac(15) ' AC cal step 15
(1A 1kHz step)
dmm.calibration.ac(16) ' AC cal step 16
(2A 1kHz step)
Example
To perform AC calibration step 1 after unlocking calibration:
dmm.calibration.ac(1)
To perform AC calibration step 2 with a value of 10 after unlocking calibration:
dmm.calibration.ac(2, 10)