Section 9
:
Instrument Control Library (ICL)
Series 3700 System Switch/Multimeter Reference Manual
9-212
Document Number: 3700S-901-01 Rev. A / August 2007
tsplink.trigger
The functions in this group control the tsp-link's trigger event detector.
tsplink.trigger[N].assert
Function
Simulates the occurrence of the trigger and generates the corresponding event id.
Usage
tsplink.trigger[N].assert()
N
: (In) The trigger line to assert (1
–3).
Remarks
This function will generate a trigger pulse on the given TSP-Link trigger line.
Also see
tsplink.trigger[N].clear
(on page 9-212)
tsplink.trigger[N].mode
(on page 9-212)
tsplink.trigger[N].overrun
(on page 9-213)
tsplink.trigger[N].release
(on page 9-214)
tsplink.trigger[N].triggersource
(see "tsplink.trigger[N].triggerstimulus" on page 9-214)
tsplink.trigger[N].wait
(on page 9-214)
tsplink.trigger[N].clear
Function
Clear the event detector for a trigger.
Usage
tsplink.trigger[N].clear()
N
: (In) The trigger line (1
–3).
Remarks
A trigger‘s event detector remembers if an event has been detected since the last
tsplink.trigger[line].wait call. This function clears a trigger‘s event detector and
discards the previous history of the trigger line.
Also see
tsplink.trigger[N].mode
(on page 9-212)
tsplink.trigger[N].overrun
(on page 9-213)
tsplink.trigger[N].release
(on page 9-214)
tsplink.trigger[N].triggersource
(see "tsplink.trigger[N].triggerstimulus" on page 9-214)
tsplink.trigger[N].wait
(on page 9-214)