Section 10
:
Verification
Series 3700 System Switch/Multimeter Reference Manual
10-14
Document Number: 3700S-901-01 Rev. A / August 2007
Applied current
Range
Current
Lower limit
Upper limit
Verify AC curr 1mA @ 1kHz
1.00E-03
1.00E-03
9.989000E-04
1.001100E-03
Verify AC curr 10mA @ 1kHz
1.00E-02
1.00E-02
9.989000E-03
1.001100E-02
Verify AC curr 100mA @ 1kHz 1.00E-01
1.00E-01
9.989000E-02
1.001100E-01
Verify AC curr 1A @ 1kHz
1.00E+00
1.00E+00
9.977000E-01
1.00
AC current verification data 3A range
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications. See
Example reading limit calculation
(on page 10-3).
Applied current
Range
Current
Lower limit
Upper limit
Verify AC curr 3A @ 1kHz
3.00E+00
3.00E+00
2.00
3.00
Verifying frequency
To verify the Series 3700 frequency function:
1. Connect the Agilent 33220A function generator to the Series 3700 INPUT pins.
2. Set the function generator to output a 1kHz, 5V RMS sine wave.
3. Select the Series 3700 frequency function by pressing the
FREQ
key.
4. Verify that the Series 3700 each frequency reading is within the limits contained in the table
contained in Frequency verification data.