Section 9
:
Instrument Control Library (ICL)
Series 3700 System Switch/Multimeter Reference Manual
9-168
Document Number: 3700S-901-01 Rev. A / August 2007
scan.state
Example
To see the current scan state:
ScanState = scan.state()
print(ScanState)
or
print(scan.state())
scan.stepcount
Attribute
Attribute to query to see number of steps in the present scan.
Usage
To read the number of steps in the present scan:
ScanStepCount = scan.stepcount
Remarks
This is a read only attribute. It is set by the number of steps in the present scan.
Also see
scan.add
(on page 9-157)
Example
To see the current scan state:
print(scan.stepcount)
scan.trigger.arm.clear
Function
Clear the arm event detector
Usage
scan.trigger.arm.clear()
Remarks
This function will set the arm event detector to the undetected state.
Example
To clear the arm event detector:
scan.trigger.arm.clear()
scan.trigger.arm.set
Function
Set the arm event detector to the detected state.
Usage
scan.trigger.arm.set()
Remarks
This function will set the arm event detector to the detected state.
Example
To set the arm event detector to the detected state:
scan.trigger.arm.set()
scan.trigger.arm.stimulus
Attribute
Arm event detector trigger selection.