Series 3700 System Switch/Multimeter Reference Manual
Section 8
:
Status Model
Document Number: 3700S-901-01 Rev. A / August 2007
8-15
Commands
Bit
To set register bits:
status.system5.enable = status.system.EXTENSION_BIT
status.system5.enable = status.system.EXT
status.system5.enable = status.system.NODEn
B0
B0
Bn
To read registers:
print(status.system5.enable)
print(status.system5.condition)
print(status.system5.event)
Refer to the following table for available n values:
Command
n value
status.system
1 to 14
status.system2
15 to 28
status.system3
29 to 42
status.system4
43 to 56
status.system5