Series 3700 System Switch/Multimeter Reference Manual
Section 9
:
Instrument Control Library (ICL)
Document Number: 3700S-901-01 Rev. A / August 2007
9-153
localnode.setup.poweron
Example
To set unit to power on with factory default settings:
localnode.setup.poweron = 0
To query power on state:
print(localnode.setup.poweron)
Output (factory default):
0
localnode.setup.recall
Function
Recalls settings from a saved setup.
Usage
localnode.setup.recall(location)
location:
Setup number to recall (0, 1, or "/usb1/<filename>").
0: Reset setup.
1: Internal setup.
<filename>:
Use the name of the desired file contained on a USB flash drive.
Remarks
If a number is sent as the parameter:
The number is interpreted as a setup number and the setup is saved to internal
memory. Setting this attribute to 0 recalls the factory default (reset) setup. Setting this
attribute to 1 recalls the user saved setup from internal memory.
If a string is sent as the parameter:
The string is interpreted as a path and file name and the setup is saved to the
corresponding file on the USB flash drive. The path may be absolute or relative to the
current working directory.
Also see
localnode.setup.save
(on page 9-154)
Example
To recall factory default settings:
localnode.setup.recall(0)
To recall the user saved setup from internal memory:
localnode.setup.recall(1)
To recall a user saved setup stored in a file named KEITHLEY_3730 on a USB flash
drive:
localnode.setup.recall("/usb1/KEITHLEY_3730")