E-2
Measurement Considerations
Model 2701 User’s Manual
Measurement considerations
Low-level voltage measurements made using the Model 2701 can be adversely affected by
various types of noise or other unwanted signals that can make it very difficult to obtain
accurate voltage readings. Some of the phenomena that can cause unwanted noise include
thermoelectric effects (thermocouple action), source resistance noise, magnetic fields, and
radio frequency interference. The following paragraphs discuss the most important of
these effects and ways to minimize them.
NOTE
For comprehensive information on low-level measurements, see the “Low Level
Measurements” handbook, which is available from Keithley.
Thermoelectric potentials
Thermoelectric potentials (thermal EMFs) are small electric potentials generated by
differences in temperature at the junction of dissimilar metals. The following paragraphs
discuss how such thermals are generated and ways to minimize their effects.
Thermoelectric coefficients
As shown in Table E-1, the magnitude of thermal EMFs generated depends on the
particular materials involved. Best results are obtained with clean copper-to-copper
connections as indicated in the table.
Table E-1
Material thermoelectric coefficients
Material
Thermoelectric potential
Copper-Copper
Copper-Silver
Copper-Gold
Copper-Cadmium/Tin
Copper-Lead/Tin
Copper-Kovar
Copper-Silicon
Copper-Copper Oxide
0.2µV/°C
0.3µV/°C
0.3µV/°C
0.3µV/°C
1–3µV/°C
40µV/°C
400µV/°C
1000µV/°C
2701-900-01.book Page 2 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...