Model 2701 User’s Manual
Basic DMM Operation
3-33
For remote programming, the following commands are valid with a 7701 module
installed:
SYSTem:FRESistance:TYPEx, NORMal
‘ Select normal 4W mode.
SYSTem:FRESistance:TYPEx, CSIDe
‘ Select common-side 4W mode.
SYSTem:FRESistance:TYPEx?
‘ Query 4W mode.
Where the x in TYPEx is the slot number for the 7701 module.
NOTE
Details on 4-wire common-side ohms measurements using the 7701 module are
provided in the manual (packing list) supplied with the module.
Temperature measurements
The Model 2701 can measure temperature using thermocouples, thermistors, and 4-wire
RTDs. When deciding which temperature sensor to use, keep in mind that the
thermocouple is the most versatile, the thermistor is the most sensitive, and the 4-wire
RTD is the most stable.
Thermocouples
For thermocouples, temperature measurement range depends on which type of
thermocouple is being used. Thermocouples that are supported include types J, K, N, T, E,
R, S, and B.
When two wires made up of dissimilar metals are joined together, a voltage is generated.
The generated voltage is a function of temperature. As temperature changes, the voltage
changes. The thermocouple voltage equates to a temperature reading. This is the basic
operation principle of the thermocouple.
NOTE
The equation to calculate thermocouple temperature is provided in Appendix F.
Type
Range
Resolution
J
-200°C to 760°C
0.001°C
K
-200°C to 1372°C 0.001°C
N
-200°C to 1300°C 0.001°C
T
-200°C to 400°C
0.001°C
E
-200°C to 1000°C 0.001°C
R
0°C to 1768°C
0.1°C
S
0°C to 1786°C
0.1°C
B
+350°C to 1820°C 0.1°C
2701-900-01.book Page 33 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...