NOVUS AUTOMATION
1/6
LogBox-RHT-LCD
TEMPERATURE AND HUMIDITY LOGGER – INSTRUCTIONS MANUAL V2.0x A
1
PRESENTATION
LogBox-RHT-LCD
is an electronic temperature and relative
humidity logger. Sensors measure these quantities and the values
obtained are visualized on the LCD display and stored in electronic
memory. The stored data is later sent to a computer, which allows it
to be viewed and analyzed in the form of tables or graphs. It is
possible to configure the humidity output to display the measured
Relative Humidity
value or the
Dew Point
temperature value.
The
NXperience
software is the tool used to set up the device and
view the data downloaded by the logger. Parameters such as
acquisition start and end times and intervals between acquisitions, for
example, can be easily configured through the
NXperience
software.
The data acquired by the logger can be exported for analysis in other
programs, such as spreadsheet programas.
1.1
IDENTIFICATION
The identification label is attached to the logger housing. Verify that
the characteristics described on this label are in accordance with
what was requested. The
RHT-LCD
model measures temperature
and relative humidity.
The following elements are shown in the logger front:
Figure 1 –
Logger Front
START / STOP button
: Button that can be configured to start or
stop acquisitions. When pressed, it also used to navigate or change
screens.
Status Indicator (LOG)
: This indicator will be active when the
logger is performing acquisitions and will be deactivated when the
acquisitions are finished or while they are not started.
Alarm Indicator (AL1 and/or AL2)
: This indicator informs you of
alarm situations that have occurred and will be active whenever an
alarm situation occurs. It will remain in this condition until a new
configuration is applied to the logger.
IR Communication Window (COM)
: This indicator is activated
when a communication attempt is made or when communication
between the logger and the computer is established.
2
SPECIFICATIONS
Operating
temperature
From -40 °C to 70 °C
Sensor
measurement
range
Temperature: -40.0 °C to 80.0 °C.
Relative Humidity (RH): 0.0 to 100.0 %,
without condensation.
Dew Point: -40.0 °C and 100.0 °C
Accuracy
See
Figure 2
.
Note
: In the
Offset
parameter of the
NXperience
software, you can erase the
measurement error found.
Measurement
resolutions
Temperature: 0.1 °C
14 bits (16383 levels)
Relative Humidity (RH): 0.1 %
12 bits (4095 levels)
Response time
Temperature: Up to 30 seconds with
fairly still air.
Humidity: Up to 8 seconds with fairly still
air (20 to 80 % RH).
Memory capacity
32.000 (32 k) logs:
Half
for each channel
or
total
when one of the channels is not
enabled.
Measurement
interval
1 second min. 18 hours max.
Power supply
3.6 Vdc lithium battery (1/2 AA), built-in
Estimated
autonomy
Higher than 200 days with weekly data
reading.
Frequent logged data readings may
shorten battery life and the interval
between acquisitions is too short.
Housing
Polycarbonate
Protection
Suitable for products with protection level
IP40
.
Dimensions
60 x 70 x 35 mm
Logger-PC data
transfer time
According to the number of logs. 40
seconds for 16.000 logs.
PC interface
Interface IR/USB or IR/Serial.
NXperience
software
Configurator software for Windows 10, 8,
7 and XP.
Menus in Portuguese, English, French or
Spanish.