LogBox-RHT-LCD
NOVUS AUTOMATION
6/6
7.5
OTHER FEATURES
Check the
NXperience
Operations Manual for more information on
the other features offered by the software, such as exporting to
other formats, filter logs, chart junction, etc.
8
TROUBLESHOOTING
Unable to communicate with the logger:
Make sure that the COM port is correctly selected and that other
software is not using that port during communication attempts.
Make sure that there is no obstacle blocking the infrared signal.
Make sure that the cable is well connected to the computer port.
Make sure that the selected port is working properly.
Logger reset:
When you are unsure about the working condition of the logger,
you can reset it. To do this, simply remove the battery and wait 2
minutes. After that, put the battery back in place.
After the reset, the logger returns to the stand-by state. If it is
performing acquisitions, it stops the process and only starts the
acquisitions when a new configuration is sent. The implemented
configuration remains unchanged after reset.
Additional information:
For further information, see the manufacturer web site.
9
SPECIAL CARE
The logger is an electronic device and some basic care is required:
•
When opening the device for battery replacement or connecting
sensors, you should avoid contact with the electronic circuitry
due to the risk of damage from static electricity.
•
Carefully check the battery polarity.
•
When closing the housing, the cover must be properly tightened
back in its place such as to assure the sealing protection.
•
Used batteries should not be recharged, dismantled or
incinerated. After use, batteries must be disposed according to
local legal rules or returned to supplier.
Note: To ensure the specified accuracy and protection level,
the transmitter must be fixed with the sensor capsule
directed downwards.
9.1
ATTENTION WITH SENSORS
If the humidity and temperature sensor is exposed to contaminating
vapors or extreme humidity and temperature conditions for extended
periods, the sensor calibration may change. To restore the
calibration, proceed as follows:
•
Remove the sensor from the capsule.
•
If solid particles are deposited on the sensor, wash it with water.
•
Place the sensor in an oven at 80 °C (+-10 °C) for 24 hours.
•
Put the sensor for 48 hours in a place with a temperature
between 20 and 30 °C and a humidity higher than 75 % RH.
•
Replace the sensor in the capsule.
IMPORTANT
The sensor from this device may be damaged or uncalibrated if
exposed to chemical agents-contaminated atmosphere. Chloridric
Acid, Nitric Acid, Sulphuric Acid and Ammonia in high
concentrations may damage the sensor. Acetone, Ethanol and
Propylene Glycol may cause a reversible measure error.
10
WARRANTY
Warranty conditions are available on our website