Close/Open Switching Module Channels
2-15
Making amps measurements
— In order to perform amps measurements, you must use
the front panel inputs of the 2701 mainframe. You can still use the non-amps module for
other aspects of the test, but you must use multiple channel operation to close channels.
Example:
NOTE
In order to use the front panel inputs, make sure the INPUT switch is in the out
(F) position.
SYST:PRES
' Restores system preset defaults.
ROUT:MULT:CLOS (@101)
' Closes channel 101.
SENS:FUNC ‘CURR:DC’
' Selects DCI function – Legal operation.
Non-measure switching modules
NOTE
Presently, non-measure Keithley modules include the Models 7705, 7711, and
7712. You can check the Keithley website (www.keithley.com) for new modules.
Keep the following in mind when using a non-measure module:
•
For a non-measure card, no channels are connected to the internal DMM (the
channels cannot be connected to the backplane).
•
Multiple channel operation should be used to close channels on a non-measure
module. For remote operation, the ROUT:MULT commands are used to close
channels.
•
Front panel system (single) channel operation cannot be used to close channels on
a non-measure module. For front panel operation, system channel operation will
cause message “NO MEAS CARD” to be displayed.
•
A non-measure module may have open/close operations that are specific only to
that module. Refer to the appropriate module manual (packing list) for details on
operation.
•
In order to perform measurements, you must use the front panel inputs of the 2701
mainframe. You can still use the non-measure module to control other operations.
2701-900-01.book Page 15 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...