![Keithley 2701 Скачать руководство пользователя страница 185](http://html2.mh-extra.com/html/keithley/2701/2701_user-manual_3950600185.webp)
5-8
Rel, Math, Ratio, Channel Average, dB
Model 2701 User’s Manual
mX+b
This math operation lets you manipulate normal display readings (X) mathematically
according to the following calculation.
Y = mX + b
where:
X is the normal display reading.
m and b are the user-entered constants for scale factor and offset.
Y is the displayed result.
NOTE
When using Rel, the rel’ed reading of the input signal is used by the mX+b
calculation.
mX+b configuration
1.
Press SHIFT and then MATH to display the math menu.
2.
Press the RANGE and key to display “mX+b” and press ENTER to display
the present scale factor:
M: +1.000000
(factory default)
3.
Key in the scale factor value. The and keys control cursor position and the
and keys increment/decrement the digit value. To change range, place the cursor
on the multiplier and use the and keys (m =
×
0.001, =
×
1, K =
×
1000, and
M =
×
1,000,000). With the cursor on the polarity sign, the and keys toggle
polarity.
4.
Press ENTER to enter the m value and display the offset (b) value:
b: +00.00000 m
(factory default)
5.
Key in the offset value.
6.
Press ENTER to enter the b value and display the one-character units designator:
UNITS: X
(factory default)
7.
Use the cursor keys and the and key if you wish to change the units designator.
The character can be any letter in the alphabet (A through Z), the degree symbol
(°), or the ohms symbol (
Ω
).
8.
Press ENTER. The MATH annunciator will turn on and the result of the calculation
will be displayed. Note that the calculation will be applied to all measurement
functions.
9.
To disable mX+b, again press SHIFT and then MATH. The MATH annunciator
will turn off.
2701-900-01.book Page 8 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...