Model 2701 User’s Manual
Remote Operations
10-31
The commands to select and configure the RS-232 interface are listed in Table 10-3.
Table 10-3
SYSTem commands to configure RS-232
Command
Description
:COMMunicate
Ethernet and RS-232 (serial) commands:
:SELect <name>
Select communications mode; SERial or ETHernet
:SERial
RS-232:
:BAUD <n>
Set baud rate; 300, 600, 1200, 2400, 4800,
9600, 19200, 38400, 57600, or 115200.
:BAUD?
Query baud rate.
:PACE <name>
Control Xon/Xoff flow control; XON (on) or
NONE (off).
:PACE?
Query state of flow control.
:TERMinator <name>
Select output terminator; CR, LF, CRLF, or LFCR.
:TERMinator?
Query output terminator.
:CONTrol
Control RTS/CTS hardware handshaking:
:RTS <name>
Enable (IBFull or RFR) or disable (OFF)
hardware handshaking.
:RTS?
Query state of hardware handshaking.
:SEND <data>
Send data (serial port) to another instrument.
:ENTer?
Read data from serial port.
2701-900-01.book Page 31 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...