Model 2701 User’s Manual
Signal Processing Sequence and Data Flow
D-3
Signal processing using instrument features
Figure D-2 shows the processing sequence for an input signal with various instrument
features enabled. If a feature is not enabled, the reading simply falls through to the next
enabled feature or to the display.
Figure D-2
Signal processing using instrument features
Input Signal
Rel
Math
(mX+b, Percent,
or Reciprocal)
Rel’ed reading = Reading - Rel value
Y = mX + b
Percent = ————––—–
X - Reference
Reference
100%
Reciprocal = 1/ X
Limits
Filter
Buffer
Display
Reading
Recall
OComp OComp reading =
Δ
V/
Δ
I
A/D
Conversion
Process
Store
Signal
Conditioning
Output trigger pulse (VMC)
Moving or Repeating
2701-900-01.book Page 3 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...