4-18
Range, Digits, Rate, Bandwidth, and Filter
Model 2701 User’s Manual
Filter control and configuration
The FILTER key toggles the state of the Filter. When the Filter is enabled, the FILT
annunciator is on. The FILT annunciator will flash when the filter is not settled. When
disabled, the FILT annunciator is off. The filter can be configured while it is enabled or
disabled.
The filter is configured from the filter configuration menu (Figure 4-4). Perform the
following steps to configure the filter:
1.
Select the desired function.
2.
Press SHIFT and then TYPE. The present WINDOW setting will be displayed.
3.
Use the RANGE or key to display the desired window setting (0.01%, 0.1%,
1%, 10%, or NONE) and press ENTER.
4.
Use the
,
, , and keys to display the number of readings to filter (1 to 100)
and press ENTER.
5.
Use the or key to display the desired filter type (moving or repeating) and
press ENTER. The filter turns on and the instrument returns to the normal
measurement state.
NOTE
While the filter is enabled (FILT annunciator on), changes to the configuration
take effect as soon as they are made. With filter disabled (FILT annunciator off),
changes to the configuration take place when the filter is enabled.
While the filtering operation is in progress, the FILT annunciator blinks. Read-
ings will continue to be processed (i.e., displayed, stored, sent over the bus), but
they could be questionable. When the FILT annunciator stops blinking, the filter
has settled.
Changing function or range causes the filter to reset. The filter then assumes the
state (enabled or disabled) and configuration for that function or range.
2701-900-01.book Page 18 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...