1-30
Getting Started
Model 2701 User’s Manual
Exercise 1 — Basic DMM measurements
The exercise in Table 1-6 measures ACV on the 10V range and stores 15 readings in the
buffer.
Table 1-6
Exercise 1—Measure AC volts - store readings in buffer
Front panel operation
Command sequence
1
For front panel operation, proceed to step 2.
For remote programming, clear the buffer
1
:
TRAC:CLE
2
Restore defaults
2
:
Press SHIFT > press SETUP > select RESTORE: FACT.
*RST
3
Select ACV function:
Press ACV.
FUNC 'VOLT:AC'
4
Select 10V range:
Press RANGE to display “RANGE: 10V”.
VOLT:AC:RANG 10
6
Store 15 readings in buffer
3
:
Press STORE > set for 000015 RDGS > press ENTER.
SAMP:COUN 15
READ?
7
Recall buffer readings
4
:
Press RECALL > use edit keys to display readings. Press EXIT to exit
recall mode.
CALC1:DATA?
1. To avoid problems with remote programming, it is good practice to routinely clear the buffer (
TRAC:CLE
) at the begin-
ning of a program that performs multiple measurements (
SAMP:COUN >1
). Restoring *RST or FACTory defaults does
not clear the buffer.
2. FACTory defaults place the instrument in a continuous measurement mode. *RST places the instrument in a non-continu-
ous measurement mode.
3. READ? triggers and returns 15 readings. These 15 readings are automatically stored in the buffer. See Exercise 4 and 5 for
more information on the READ? command.
4. Statistics for buffer readings are also stored in the buffer. For remote programming,
CALC1:DATA?
only returns the read-
ings that were stored. It does not return buffer statistics.
CALC2
commands are used to calculate and return buffer statistics
(see Section 6 for details).
2701-900-01.book Page 30 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...