![Keithley 2701 Скачать руководство пользователя страница 310](http://html2.mh-extra.com/html/keithley/2701/2701_user-manual_3950600310.webp)
Model 2701 User’s Manual
Remote Operations
10-23
Query commands
This type of command requests (queries) the presently programmed status. It is identified
by the question mark (?) at the end of the fundamental form of the command. Most
commands have a query form:
TRIGger:TIMer?
Queries the timer interval.
Most commands that require a numeric parameter (<n>) can also use the DEFault,
MINimum, and MAXimum parameters for the query form. These query forms are used to
determine the *RST default value and the upper and lower limits for the fundamental
command. Examples are:
TRIGger:TIMer? DEFault
Queries the *RST default value.
TRIGger:TIMer? MINimum
Queries the lowest allowable value.
TRIGger:TIMer? MAXimum
Queries the largest allowable value.
Case sensitivity
Common commands and SCPI commands are not case sensitive. You can use upper or
lower case and any case combination. Examples:
*RST
= *rst
DATA?
= data?
SYSTem:PRESet = system:preset
Long-form and short-form versions
A SCPI command word can be sent in its long-form or short-form version. The command
subsystem tables in Section 15 provide the long-form version. However, the short-form
version is indicated by upper case characters. Examples:
SYSTem:PRESet
long-form
SYST:PRES
short-form
SYSTem:PRES
long-form and short-form combination
Note that each command word must be in long-form or short-form and not something in
between. For example, SYSTe:PRESe is illegal and will generate an error. The command
will not be executed.
2701-900-01.book Page 23 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...