Model 2701 User’s Manual
Signal Processing Sequence and Data Flow
D-7
Data flow (remote operation)
Remote operation can be used with triggering configured to perform a specified number of
measurements and then stop. The various read commands (SENS:DATA?, FETCh?,
READ?, MEAS?, CALC2:DATA?, TRACe:DATA?, and CALC1:DATA?) return the data
array(s) acquired during the measurement cycle. Data flow for this triggering
configuration is summarized by the block diagram shown in Figure D-4. Refer to this
block diagram for the following discussion.
Figure D-4
Data flow for remote operation
SENSe
Measurement, Filter,
Rel, and Ratio
or Ch Avg
CALC1
CALC3
Limit Tests
TRACe:DATA?
CALC2
Min, Max, Sdev
Mean, Pk-Pk
[SENS[1]]:DATA[:LAT]?
[SENS[1]]:DATA:FRESh?
Math (mX+B,
Percent,
or 1/X)
CALC3:LIM1:FAIL?
CALC3:LIM2:FAIL?
FETCh?
READ?
MEAS?
CALC[1]:DATA[:LAT]?
CALC[1]:DATA:FRESh?
Sample
Buffer
TRACe
Data Store
TRAC:CLE
INIT:CONT OFF
TRIG:COUN 1
INIT
CALC2:IMM?
CALC2:IMM
CALC2:DATA?
2701-900-01.book Page 7 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...