![Keithley 2701 Скачать руководство пользователя страница 162](http://html2.mh-extra.com/html/keithley/2701/2701_user-manual_3950600162.webp)
Model 2701 User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-7
Setting digits
Even though the parameters for the DIGits command are expressed as integers (4 to 7),
you can specify resolution using a real number. For example, to select 3½ digit resolution,
let <n> = 3.5. Internally the instrument rounds the entered parameter value to the nearest
integer.
As implied by the commands in Table 4-3, each mainframe input function can have its
own unique digits setting.
Digits programming examples
NOTE
The following examples can be run from the KE2700 Instrument Driver using
the example named “Digits” in Table G-1 of Appendix G.
Example #1 — The following commands set digits for DCV and ACI:
VOLT:DIG 5.5
' Set DCV to 5
½
digits.
CURR:AC:DIG 4
' Set ACI to 3
½
digits.
Example #2 — The following command sequence configures channels 101 through 110 of
the Model 7700 to select 4½ digits when they are scanned.
FUNC 'RES', (@101:110)
' Select
Ω
2 function.
RES:DIG 4.5,(@101:110)
' Set scan channels to 4
½
digits.
Table 4-3
Digits commands
Commands*
Description
Default
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:DIGits <n> [, <clist>]
Set # of digits for DCV; <n> = 4 to 7.
7
:VOLTage:AC:DIGits <n> [, <clist>]
Set # of digits for ACV; <n> = 4 to 7.
6
:CURRent[:DC]:DIGits <n> [, <clist>]
Set # of digits for DC1; <n> = 4 to 7.
7
:CURRent:AC:DIGits <n> [, <clist>]
Set # of digits for ACI; <n> = 4 to 7.
6
:RESistance:DIGits <n> [, <clist>]
Set # of digits for
Ω
2; <n> = 4 to 7.
7
:FRESistance:DIGits <n> [, <clist>]
Set # of digits for
Ω
4; <n> = 4 to 7.
7
:TEMPerature:DIGits <n> [, <clist>]
Set # of digits for TEMP; <n> = 4 to 7.
6
:FREQuency:DIGits <n> [, <clist>]
Set # of digits for FREQ; <n> = 4 to 7.
7
:PERiod:DIGits <n> [, <clist>]
Set # of digits for PERIOD; <n> = 4 to 7.
7
Channel list parameter:
<clist> = (@SCH)
where: S =
Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
* The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the digits command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:DIGits 4.5, (@101) is only
valid if scan channel 101 is set for the ACV function.
2701-900-01.book Page 7 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...