Model 2701 User’s Manual
Limits and Digital I/O
9-17
Digital outputs
With the digital outputs of the Model 2701 enabled, the digital outputs will respond as
follows for each resistor reading:
LO limit 2
LO limit 1
HI limit 1
HI limit 2
Resistor
tolerance
bin
Affected
outputs*
Pass
Pass
Pass
Pass
1%
1
None
Pass
Fail
Pass
Pass
5%
2
#1 and #5
Pass
Pass
Fail
Pass
5%
2
#2 and #5
Fail
Fail
Pass
Pass
>5%
3
#1, #3, and #5
Pass
Pass
Fail
Fail
>5%
3
#2, #4, and #5
*Affected outputs are pulled (or pulsed) high or low when a limit test fails.