3-6
Basic DMM Operation
Model 2701 User’s Manual
Remote programming — autozero and LSYNC
Autozero and LSYNC commands
The commands to control autozero and line synchronization are listed in Table 3-1.
Table 3-1
Autozero and LSYNC commands
Commands
Description
Default
Autozero command
*
SYSTem:AZERo[:STATe] <b>
Enable or disable autozero; <b> = ON or
OFF
ON
Line synchronization command
SYSTem:LSYNc[:STATe] <b>
Enable or disable LSYNC; <b> = ON or
OFF
OFF
* After enabling autozero, you can update the internal reference points immediately by setting the integration
rate to 0.002 PLC and then back to the desired setting (see NPLC commands in Section 4).
2701-900-01.book Page 6 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...