Intellipuff Service Manual
Page 36 of 51
Issue D 27/02/2014
EP59-12317
The mainframe and PCB assembly including the valve coil is now free. Take
care that the coil assembly does not damage the cables to it as the mainframe
and PCB assembly is handled.
Note that further disassembly of this unit is NOT recommended. The unit is
set up and tested as a unit. Adjustment of the unit is also discouraged as it
has been set up to a strict test procedure using jigs only available in our
production department.
Replace Mainframe and PCB Assembly
Lay the case half with the browrest fitted down.
Refit the umbilical cable with the 12 way connector onto the PCB
Lay the mainframe and PCB assembly onto the case half carefully following
the routing of the cables as shown below.
Flatten the cables on top of the mainframe so that they do not cross each
other and retain them in place with a piece of insulating tape.