Intellipuff Service Manual
Page 22 of 51
Issue D 27/02/2014
EP59-12317
Selecting the Print radio button, enables the operator to print the current
calibration parameters along with the table of ADC/Pressure values by
pressing the Print button. This calibration report also includes unit serial
number for future reference. If the user selects to finish programming the user
will be prompted to confirm that the calibration run is not required. After
completion of this form, the user can press the Cancel button and the Mode
selection form is loaded, from which the Exit button is pressed to return the
unit under test to local mode. At this point the unit will switch itself off to reset
the system.
Reading the EEPROM Data
The Mode selection form also provides a button for reading the calibration
constants stored in the EEPROM of the unit currently under test as both raw
data and scaled back to their natural units where appropriate. These data are