P/N 960-100929R_Rev. 7 {EDP #148849}
© 2017, JAPAN CASH MACHINE CO., LTD.
6 - 1 0
Section 6
iVIZION® SeriesNext-Generation Banknote Acceptor Unit
Calibration and Testing
20. When the Calibration Value writing EEPROM
writing is complete, the “
Calibration Informa-
tion
” Screen shown in Figure 6-46 will appear.
21. Click on the “
OK
” Screen Button (Figure 6-46
a
).
This completes the Paper Calibration Procedures.
Refer to “Reference Paper Use Precautions” on
page 1-4 to properly store and prevent damage to
the KS-072/KS-089 Reference Paper.
Calibration Plus Serial Number Writing
When the Sensor Calibration procedure is com-
plete, the capability to write the Serial Number into
the iVIZION Unit becomes possible. To write a
Serial Number into Memory, proceed as follows:
1. Complete the initial Sensor Calibration Procedure
first (See “Calibration Only” on page 6-7 of this
Section).
2. Once the CIS Sensor Calibration is complete, the
“
iVIZION Calibration Service Suite Edition
”
Screen shown in Figure 6-47 will appear.
3. Click on the “
OK
a
).
The “
Setting Manufacture No
” Screen shown in
Figure 6-48 will then appear.
4. Click on either the “
▲
” or “
▼
” Triangle Indica-
tion Buttons in each Serial Number Box located
at the bottom of the Screen to either increase or
decrease the numbers to change (Figure 6-48
a
).
5. Click on the “
OK
” Screen Button located on the
right side of the Screen (Figure 6-48
b
).
The Calibration Value and the Serial Number will
then be written into the EEPROMs on the Sensor
and the CPU Boards (Figure 6-49).
When the data writing into the EEPROM is com-
plete, the “
Calibration Information
” Screen
shown in Figure 6-50 will appear.
6. Click on the “
OK
” Screen Button (Figure 6-50
a
).
This completes the Serial Number Writing
Procedure.
Figure 6-46
Figure 6-46
Calibration Complete Screen
a
NOTE: To perform the “Writing Serial
No.”, process, click on the “Update
serial No.” Check-box when the
“iVIZION Calibration Ver.X.XX” Screen
is active (See iVIZION Calibration
b
on
page 6-7 of this Section).
Figure 6-47
iVIZION Calibration Service Suite
Figure 6-47
iVIZION Calibration Service Suite
Edition Screen
a
Setting Manufacture No. Screen
Figure 6-48
Setting Manufacture No. Screen
a
b
Calibration Value Writing Screen
Figure 6-49
Calibration Value Writing Screen
NOTE: The Barograph does not indicate a Calibration in progress.
Figure 6-50
Figure 6-50
Adjustment Info Screen
a