Calibration and Testing
iVIZION® SeriesNext-Generation Banknote Acceptor Unit
Section 6
13. Remove the Cash Box and remove the piece of
Cardboard, then reseat the Cash Box into the
iVIZION Unit again.
14. Place Reference Paper KS-072/KS-089 into the
iVIZION Unit (See “Placing the KS-072/KS-089
Reference Paper” on page 6-6 of this Section).
15. Click on the “
OK
” Screen Button (Figure 6-39
a
)
to begin the following two (2) UV Sensor Cali-
bration Procedures:
-
Lower UV Sensor
-
Upper UV Sensor
Calibration will be performed in the above alpha-
betic order while “
UV
RUNNING
” is showing on
Once the UV Sensor Calibration is complete, the
“
Calibration Information
” Screen shown in Fig-
ure 6-41 will appear.
16. Remove the KS-072/KS-089 Reference Paper and
click on the “
OK
” Screen Button (Figure 6-41
a
)
to begin the UV Sensor Non-Paper Calibration.
Once the UV Sensor Non-Paper Calibration is
complete, the “
Calibration Information
” Screen
shown in Figure 6-42 will appear.
17. Place the KS-072/KS-089 Reference Paper back
into the iVIZION Unit (See “Placing the KS-072/
KS-089 Reference Paper” on page 6-6 of this
Section).
18. Click on the “
OK
” Screen Button (Figure 6-42
a
)
to begin the following three (3) CIS Calibration
Procedures:
-
CIS Transmissive
-
CIS Upper
-
CIS Lower
Calibration will be performed in the above order
while “
CIS
RUNNING
” is showing on the
Once the CIS Calibration is complete, the “
Cali-
bration Information
” Screen shown in Figure 6-
44 will re-appear.
19. Remove the KS-072/KS-089 Reference Paper and
click on the “
OK
” Screen Button (Figure 6-44
a
)
to begin the CIS Sensor Non-Paper Calibration
Procedure.
Once the CIS Non-Paper Calibration Procedure is
complete, the Calibration Value will be written to
the EEPROMs located on the Sensor Board and
Calibration Information Screen 3
Figure 6-39
Calibration Information Screen 3
a
Figure 6-40
UV Calibration Proceeding Screen
Figure 6-40
UV Calibration Proceeding Screen
NOTE: The Barograph does not indicate a Calibration in progress.
Figure 6-41
Calibration Information Screen 4
Figure 6-41
Calibration Information Screen 4
a
Figure 6-42
Calibration Information Screen 5
Figure 6-42
Calibration Information Screen 5
a
Figure 6-43
Figure 6-43
CIS Calibration Proceeding Screen
NOTE: The Barograph does not indicate a Calibration in progress.
Figure 6-44
Calibration Information Screen 6
Figure 6-44
Calibration Information Screen 6
a
Figure 6-45
Calibration Value Writing Screen
Figure 6-45
Calibration Value Writing Screen
NOTE: The Barograph does not indicate a Calibration in progress.