PR-655/670 User Manual
46
7.
Measure the sample with the
Target
instrument.
8.
The
Custom Spectral Calibration
is complete.
9.
To use these new factors for subsequent measurements, make sure that
Custom Spectral Correction
is
Enabled
and that
One Time Calibration
is
Disabled
.
10.
To restore the Target instrument so that it uses only the factory calibration factors, make sure
Custom Spectral Correction
is set to
Disabled
.
A
S A
.TXT
F
ILE
Data from other instruments can be used as the baseline or
Source
values for the Custom Spectral Calibration.
The following criteria must be met:
A.
The
Source
must have the same starting and ending wavelength points, and the same wavelength
(data) increment.
B.
The file must have a
.TXT
file extension.
C.
The file must have the following file format:
Line 1
–
Title (Alpha / Numeric characters only
–
no spaces)
Line 2
–
380 (Starting Wavelength)
Line 3
–
780 - Ending Wavelength
Line 4
–
4 (for PR-655) or 2 (for PR-670)
–
Wavelength (Data) Increment (in nanometers)
Line 5 to Line N
–
Spectral values
Sample
.txt
File:
M1
3.002
7.002
1.000
3.722937e-003
3.729762e-003
3.611207e-003
3.971064e-003
4.476557e-003
4.434512e-003
4.972667e-003
4.636224e-003
5.055876e-003
.
.
.
.
.
.
1.
Create and save the
Source
spectral values on the SD card. Make sure this file does not have the same name
as another file on the same card with a
.txt
extension.
2.
Insert the SD card in the
Target
instrument.
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...