PR-655/670 User Manual
133
R
EMOTE
C
ONTROL
E
RROR
C
ODES
R
EMOTE
C
ONTROL
M
EASUREMENT
E
RRORS
Error
Meaning
-1
Light source not constant.
-2
Light overload
–
signal too intense.
-3
Cannot Sync to light source. Light source frequency below 20Hz, above 400
Hz or signal too low to Sync.
-4
Adaptive mode error.
-8
Weak light
–
insufficient signal.
-9
Sync Error.
-10
Cannot Auto Sync to light source.
-12
Adaptive mode time out. Light source not constant.
R
EMOTE
C
ONTROL
P
ARSING
E
RRORS
Error
Meaning
Valid Values
-1000
Illegal command
-1001
Too many fields in setup command
-1002
Invalid primary accessory code
-1003
Invalid Addon 1 accessory code
-1004
Invalid Addon 2 accessory code
-1005
Accessory is not a primary accessory
-1006
Accessory is not an Addon accessory
-1007
Accessory already selected
-1008
Invalid Aperture index (PR-670 only)
-1009
Invalid units code
0 = English
1 = Metric (SI)
-1010
Invalid Exposure value
PR-655
3 to 6000 ms
PR-670
6 to 30,000 ms
-1011
Invalid Gain code
0 = Normal
1 = 1X for AC sources
2 = 10X
3 = 100X
-1012
Invalid average cycles
1 to 99
Содержание PHOTO RESEARCH SpectraScan PR-655
Страница 2: ...This page intentionally left blank...
Страница 25: ...PR 655 670 User Manual 9 FIGURE 10 SAMPLE EXTERNAL TRIGGER INPUT CIRCUIT DRIVERS...
Страница 106: ...PR 655 670 User Manual 90 FIGURE 87 USB POWER SCREEN Total power available Total power required...
Страница 159: ...PR 655 670 User Manual 143 PR 655 PR 670 OUTLINE DRAWINGS...
Страница 160: ...PR 655 670 User Manual 144...